Cyrogenic air separation is used to produce high-purity oxygen, nitrogen, Argon and carbon dioxide. Highly accurate oxygen measurements are a crucial part of the quality assurance process. The XZR400 uses Michell’s MSRS zirconium oxide oxygen sensor, which contains a metallic sealed reference and ensures long-term repeatability of measurements as no reference air is required. Unlike electro-chemical sensors, which require regular replacement, the MSRS sensor of the XZR400 should last in excess of 7 years. Calibration is not needed as frequently either (every 3 – 6 months) and can be carried out with just one calibration gas saving time and money.
The XZR400 series oxygen analyzers offer three configurations: rack, wall and bench mount. All models include an intuitive touch-screen interface for quick and easy operation, barometric pressure, digital flow meters and a 4-20mA output as standard. Options include additional outputs such as Modbus RTU over RS485 protocol, RS232 and process pressure correction.